REMOTE CONTROL

123

Measuring Equipment Configuration Commands

EQUipment

Sets or queries the EUT class.

Syntax

EQUipment {CLAss1|CLAss2|INTernal}

Query Syntax

EQUipment?

Query Return

Returns the equipment class of the EUT:
CLAss1|CLAss2|INTernal

CLAss1: CLASS I
CLAss2: CLASS II

INTernal:Internally
powered

Example

EQUipment CLAss1
Sets the class of the EUT to “CLASS I”.

Query Example

EQUipment?
Return: CLASS1
Returns the current EUT class.

EQUipment:TYPE

Sets or queries the applied part of the EUT.
Note: This command can only be used with medical networks/Network B

Syntax

EQUipment:TYPE {B|BF|CF}

Query Syntax

EQUipment:TYPE?

Query Return

Returns the applied part of the EUT: B|BF|CF (Medical
Network only)

B: B Type

BF: BF Type

CF: CF Type

Example

EQUipment:TYPE BF
The medical network is set to type BF.

Query Example

EQUipment:TYPE?
Return: BF
Type BF is the currently applied part of the EUT for
network B/Medical networks.