Class I/Class II

1. Position the T2 test lead to the applied part of

the EUT.

Class I and Class II

Measurement
Setup

Confirm all settings including MD, leakage current
mode, measurement time, upper and lower limits,
and other parameters.

Panel Operation

2. Turn on the circuit breaker.

Connect the EUT to the
power socket.



GLC-9000

3. Press the START button to

start measurements.



GLC-9000

4. Press the reset button to

stop measurement.



GLC-9000

Warning

Turn the circuit breaker off before removing the EUT.

Ensure the power consumption of the EUT doesn’t
exceed the rated power limits.