REMOTE CONTROL

125

Query Example

MODE?
Return: EARTH
Earth leakage current is the current measurement
mode.

Measurement Commands

CONFigure:AUTO

Configures or queries the current measurement mode. When the
measurement mode is set to Auto, a measurement can be started after
any of the following:
The

Start

key on the panel has been pressed.

The START line of the EXT I/O line is sent low
The STARt command is used.

Syntax

CONFigure:AUTO {ON|OFF}

Query Syntax

CONFigure:AUTO?

Query Return

Returns the current measurement mode as a string:
(ON|OFF)

ON: Automatic mode

OFF: Manual mode

Example

CONFigure:AUTO OFF
Sets the measurement mode to manual mode.

Query Example

CONFigure:AUTO?

Return: OFF
The current measurement mode is manual.

CONFigure:FILTer

Sets or queries whether Network F has the RC network enabled/disabled.
If the RC network is OFF, a 1k pure resistance filter is used.

Syntax

CONFigure:FILTer {ON|OFF}

Query Syntax

CONFigure: FILTer?

Query Return

Returns a string indicating if the RC network is ON or
OFF.

ON: RC filter is ON.