Enclosure to Enclosure Leakage Current

Network

Non medical network / Medical network

Connection

As illustrated, ensure that the power source and
test leads are properly connected.

Non-Medical
network

1. Connect the tests lead to terminals T1& T2.

2. Position the test leads on un-grounded sections

of the enclosure.

Non-medical type (General electrical equipment)

Medical Network

1. Connect the test leads to the T1 & T2 terminals

and position the test leads on ungrounded
sections of the enclosure.

2. Connect a test lead to the T3 terminal and

position the test lead on an ungrounded section
of signal I/O on the EUT.