OVERVIEW

15

5. Patient

Leakage
Current III

Refers to the current that flows through a human
body that is directly connected to a device that
malfunctions.

(Medical Equipment)

6. Patient

Auxiliary
Current

Refers to the current that flows through a human
and a device during normal operation. This is not
intended to be perceptive. (Medical Equipment)

Measurement Principles

Background

Leakage current can be categorized into 3 types:
Electric current that flows through the body of
someone who touches the equipment, the current that
flows through a protective grounding wire to earth and
the current that flows through a human body
connected to an applied part. When testing the
leakage current of an EUT (equipment under test),
testing must be performed under normal and single
fault conditions.

As shown below, a number of relays are used to
simulate different fault conditions. Power to the EUT
is normally open and thus turning off the power
disconnects one wire on the power line.