Measurement Mode Command

MODE

Set or queries the measurement mode.
Note: Different measuring networks have different measurement modes.

Syntax

MODE
EARTh|ENCLosure1(ENCL1)|ENCLosure2(ENCL2)
|ENCLosure3(ENCL3)|PATient1(PAT1)|PATient2(PAT2)
|PATient3(PAT3)|PAUXiliary(PAUX)

Query Syntax

MODE?

Query Return

Returns the measurement mode:
EARTH|ENCLOSURE1|ENCLOSURE2|ENCLOSURE3|P
ATIENT1|PATIENT2|PATIENT3|PAUXILIARY

When measurement network A,B,C,D,E,G,H or I is
selected:

EARTh

Earth leakage current.

ENCLosure1

Enclosure to earth leakage current.

ENCLosure2

Enclosure to enclosure leakage
current.

ENCLosure3

Enclosure to line leakage current.

When measurement network F is selected:

EARTh

Earth leakage current

ENCLosure1

Enclosure to earth leakage current

ENCLosure2

Enclosure to enclosure leakage
current

PATient1

Patient leakage current I

PATient2

Patient leakage current II

PATient3

Patient leakage current III

PAUXiliary

Patient Auxiliary Current

Example

MODE EARTH
Sets the measurement mode to Earth leakage current.