Enclosure Leakage Current

Network

Non medical network / Medical network

Connection

As illustrated, ensure that the power source and
test leads are properly connected.

Non-Medical
Network

1. Connect the test lead to terminal T2.

2. Position the test lead on an ungrounded section

of the enclosure

Non-medical type (General electrical equipment)

Medical Network

1. Connect a test lead to the T2 terminal and

position the test lead on an ungrounded section
of the enclosure.

2. Connect a test lead to the T3 terminal and

position the test lead on an ungrounded section
of signal I/O on the EUT.