Patient leakage current iii – GW Instek GLC-9000 User Manual User Manual

Page 56

background image

GLC-9000 User Manual

56

4. Press the START button to

start measurements.



GLC-9000

5. Press the reset button to

stop measurement.



GLC-9000

Warning

Turn the circuit breaker off before removing the EUT.

Ensure the power consumption of the EUT doesn’t
exceed the rated power limits.

Patient Leakage Current III

Network

For medical network circuits only (MD-F), type F
only.

Connection

As illustrated, ensure that the power source and
test leads are properly connected.

Type F Medical
Network

1. Connect the test lead to the T2 terminal and

position the test lead to the applied part of the
EUT.