beautypg.com

GW Instek GLC-9000 User Manual User Manual

Page 133

background image

REMOTE CONTROL

133

Machine Status

Int power

Normal

Single Fault Condition

Measurement
Mode

Power Line

Disconnected

Earth

Disconnected

110%

output

positive

polarity

110%

output

negative

polarity

Earth Leakage
Current

--

--

--

--

--

Enclosure and
earth leakage
current

--

--

Enclosure to
Enclosure
leakage current

--

--

Patient Auxiliary
current

--

--

--

--

Patient Leakage
Current I

--

--

--

--

Patient Leakage
Current II

--

--

--

Patient Leakage
Current III

--

--

--

●Selectable, -- Not selectable

CONFigure:POLarity

Sets or queries the power supply polarity in manual testing.
Note: The power supply polarity can only be set in manual
measurement/mode.

Syntax

CONFigure:POLarity {NORMal|REVerse}

Query Syntax

CONFigure:POLarity?

Query Return

Return value:
Returns the polarity of the power supply in manual
measurement
NORMal|REVerse

NORMal

positive polarity

REVerse

negative polarity