Leakage current modes – GW Instek GLC-9000 User Manual User Manual

Page 29

background image

OVERVIEW

29

Leakage Current Modes

Type

Normal
Condition

Single
Fault

Fault Description

Notes

Earth Leakage
Current

Yes

Yes

Power line disconnected.

Class 1
only

Enclosure to
Earth leakage
current

Yes

Yes

1. Power line

disconnected.

2. Disconnected earth

protection*.

3.110% of the rated power

supply voltage applied
to the signal
input/output parts and
earth.


Class 1
only*

Enclosure to
Enclosure
leakage current

Yes

Yes

Enclosure to
line leakage
current.

Yes

Yes

Patient Auxiliary
Current

Yes

Yes

1. Power Line

disconnected.

2. Earth disconnected*.

Class 1
only*

Patient Leakage
Current I

Yes

Yes

Patient Leakage
Current II

No

Yes

110% of the rated
power supply voltage
applied to the signal
input/output parts and
earth.

Type B
applied
part

Patient Leakage
Current III

No

Yes

110% of the rated
power supply voltage
is applied to the F-type
applied part and earth.

Type F

applied
part