GLC-9000 User Manual

106

Zero Adjustment

Zero Adjustment is used to compensate for
extension or adapter cables that are connected to
an EUT power cable for Earth (Ground) leakage
tests. Zero adjustment is only applicable in AC or
AC Peak modes in earth leakage mode.

Conditions

Class I
Earth Leak mode
AC or AC Peak
current

Page 63
Page 67
Page 72

1. Ensure the EUT cable is disconnected from the

extension or adapter cable that is connected to
the leakage current tester.

2. To perform a zero

adjustment, press Zero.

Zero

3. Chose Yes to accept or No to

cancel the Zero adjustment at
the next screen.

Yes

OR

No