REMOTE CONTROL

135

NLINe

Application of voltage from the
neutral line. Normal neutral line
connection.

Example

CONFigure: AMITem <101>

256

128

64

32

16

8

4

2

1

Bit8

bit7

bit6

bit5

bit4

bit3

bit2

bit1

bit0

0

0

1

1

0

0

1

0

1

Automatic measurement items include: normal power
supply, disconnected earth line, normal polarity, and
reverse polarity.

Query Example

CONFigure: AMITem?
Return: 4
Power source disconnected is the current
measurement setting.

CONFigure:AMTime

Set or query the auto measurement duration, ranging from 1~300
seconds.
Note: Auto measurement configuration cannot be used in manual mode.

Syntax

CONFigure:AMTime

Query Syntax

CONFigure:AMTime?

Query Return

Returns the auto measurement time value.

1~300 secs.

Example

CONFigure:AMTime 2
Set the duration of the auto measurement to 2 secs.

Query Example

CONFigure:AMTime?
2
Returns the auto measurement time (2 seconds).