System setup command – GW Instek GLC-9000 User Manual User Manual

Page 141

background image

REMOTE CONTROL

141

1.040mA

Positive

Power line

disconnected

DC

FAIL-U

1.019mA

Negative

Power line

disconnected

DC

PASS

1.013mA

Positive

Earth

disconnected

DC

FAIL-L

1.027mA

Negative

Earth

disconnected

DC

PASS

Note

If there’s no return value for the file no. query, 0 will be
returned.

MEMory:SAVE:AUTO

Saves auto measurement results, including file name, file no.,
instrumentation class, medical network application, network,
measurement mode, measurement type, leakage current, leakage current
range, maximum limit, minimum limit, measuring conditions, power
supply polarity and auto measurement items.

Syntax

MEMory:SAVE:AUTO

Example

MEMory:SAVE:AUTO
Saves auto measurement results.

System Setup Command

SYStem:BACKlight

Sets or queries the backlight to turn off automatically after a set time.

Syntax

SYStem:BACKlight

Range: 0~30

Query Syntax

SYStem:BACKlight?

Query Return

Returns the wait time (in minutes) before the
backlight will automatically turn off.

0

Infinite (backlight will not
automatically turn off).

1~30

Time in minutes before the
backlight will turn off.