EXTERNAL I/O

167

16

Input

LOAD2

Selects a panel setting to load.
LOAD2 is bit 3 of 5

17

Input

LOAD4

Selects a panel setting to load.
LOAD5 is bit 5 of 5

18

---

Reserved

19

Output

MEAS

The MEAS signal goes low for each
measurement item during automatic
measurement.

20

Output

H-FAIL

Active on a FAIL judgement
(exceeding upper limit)

21

---

Reserved

22

Input

VDC-EXT Power supply input from external

equipment: 5~24V DC

23

Input

VDC-EXT

24

Input

GND-EXT Ground input from external

equipment

25

Input

GND-EXT

LOAD0~LOAD4 control table and corresponding panel settings

Panel no.

LOAD4

LOAD3

LOAD2

LOAD1

LOAD0

1

1

1

1

1

0

2

1

1

1

0

1

3

1

1

1

0

0

4

1

1

0

1

1

5

1

1

0

1

0

6

1

1

0

0

1

7

1

1

0

0

0

8

1

0

1

1

1

9

1

0

1

1

0

10

1

0

1

0

1

11

1

0

1

0

0

12

1

0

0

1

1

13

1

0

0

1

0