Medical equipment, Earth leakage current – GW Instek GLC-9000 User Manual User Manual

Page 44

background image

GLC-9000 User Manual

44

Medical Equipment

MD-F

CLASS I

CLASS II

Internal Power

Type B Type F* Type B

Type F* Type B Type F*

Earth leakage current

Not Used

N/A

N/A

Enclosure and
earth leakage
current

Normal

T2

T2

T2

T2

T2

T2

Fault

T2,T3

T2,T3

T2,T3

T2,T3

T2,T3

T2,T3

Enclosure and
enclosure leakage
current

Normal

T1,T2

T1,T2

T1,T2

T1,T2

T1,T2

T1,T2

Fault

T1~3

T1~3

T1~3

T1~3

T1~3

T1~3

Patient auxiliary current

T1,T2

T1,T2

T1,T2

T1,T2

T1,T2

T1,T2

Patient leakage current I

T2

T2

T2

T2

T1,T2

T1,T2

Patient leakage current II

T2,T3

N/A

T2,T3

N/A

T2,T3

N/A

Patient leakage current III

N/A

T2

N/A

T2

N/A

T2

*Type F = Type BF and Type CF.

Earth Leakage Current

Network

N/A.

Connection

1. Connect the EUT power cord to the GLC-9000

as shown in the diagram below.

Measurement
Setup

Confirm all settings including MD, leakage current
mode, measurement time, upper and lower limits,
and other parameters.