Specifications, Appendix, And ½ respectively. dc – GW Instek GLC-9000 User Manual User Manual
Page 177: Ac/ ac+dc, Ac peak
APPENDIX
177
Specifications
Specification accuracy is only applicable when the GLC-9000 has
been warmed up for 30 minutes and has an operating temperature
of +18°C – +28°C. The specifications below are based on a 1kΩ
purely resistive network. For networks G and H scale the range by
1/1.5
and ½ respectively.
DC
Ranges
Range
Resolution
Accuracy
25.00mA
5.00mA~25.00mA
10uA
±(0.2%rdg+3dgt)
5.000mA
0.500mA~5.000mA
1uA
±(0.2%rdg+3dgt)
500.0uA
50.0uA~500.0uA
0.1uA
±1.0%fs
50.00uA
4.00uA~50.00uA
0.01uA
±1.0%fs
AC/ AC+DC
Ranges
Range
Resolution
Accuracy
10Hz≦f≦100kHz 100kHz 25.00mA 5.00mA~25.00mA 10uA ±(2.0%rdg+6dgt) ±(2.0%rdg+10dgt) 5.000mA 0.500mA~5.000mA 1uA ±(2.0%rdg+6dgt) ±(2.0%rdg+10dgt) 500.0uA 50.0uA~500.0uA 0.1uA ±(2.0%rdg+6dgt) ±(2.0%rdg+10dgt) 50.00uA 4.00uA~50.00uA 0.01uA ±2.0%fs ±2.0%fs AC Peak Ranges Range Resolution Accuracy 20Hz≦f≦1kHz 1kHz 75.0mA 10.0mA~75.0mA 100uA ±(2.0%rdg+2dgt) ±(5.0%rdg+10dgt) 10.00mA 1.00mA~10.00mA 10uA ±(2.0%rdg+2dgt) ±(5.0%rdg+10dgt) 1.000mA 0.500mA~1.000mA 1uA ±2.5%fs ±5.0%fs 500.0uA 40.00uA~500.0uA 0.1uA ±4.0%fs ±5.0%fs EUT Voltage / Current Range Resolution Accuracy 300V 85V~300V 0.1V ±(2%rdg+10dgt) 10A 0.5A~10A 0.1A ±(2%rdg+5dgt) Voltage Range Resolution Accuracy AC 10.0~300.0V 0.1V ±(3%rdg+2V) DC 10.0~300.0V 0.1V ±(3%rdg+2V) AC+DC 10.0~300.0V 0.1V ±(3%rdg+2V) AC Peck 15.0~430.0V 0.1V ±(3%rdg+2V)
Ranges
Ranges