GW Instek GLC-9000 User Manual User Manual

Page 140

background image

GLC-9000 User Manual

140

3: Normal neutral line connection
conditions, 110% voltage
application.
4: Negative phase, 110% voltage
application.
5: Normal live line connection
conditions. Application of voltage
from the live line.
6:Normal neutral line connection
conditions. Application of voltage
from the neutral line.

Leakage current type:
0: AC
1: DC
2: AC+DC
3: ACpeak

Judgment Value:
0: Measurement is within upper
and lower judgment limits (PASS)
1: Measurement is greater than the
upper limit (FAIL-U)
2: Measurement is less than the
lower limit (FAIL-L)

Query Example

MEASure:AUTO?6
file_01
+1.031E-03,0,0,1,0,+1.024E-03,1,0,1,0
+1.040E-03,0,1,1,1,+1.019E-03,1,1,1,0
+1.013E-03,0,2,1,2,+1.027E-03,1,2,1,0

Return value examples are shown in the table below:

Maximum

Power

polarity

Equipment

Status

Leakage

current type

Judgment

1.031mA

Positive

Normal

DC

PASS

1.024mA

Negative

Normal

DC

PASS