Save data command – GW Instek GLC-9000 User Manual User Manual

Page 138

background image

GLC-9000 User Manual

138

Judgment Value:
0: Measurement is within upper
and lower judgment limits (PASS)
1: Measurement is greater than the
upper limit (FAIL-U)
2: Measurement is less than the
lower limit (FAIL-L)

Query Example

MEASure:AUTO?
File_01
+1.031E-03,0,0,1,0,+1.024E-03,1,0,1,0
+1.040E-03,0,1,1,1,+1.019E-03,1,1,1,0
+1.013E-03,0,2,1,2,+1.027E-03,1,2,1,0

Return value examples are shown in the table below:

Maximum

Power

polarity

Equipment

Status

Leakage

current type

Judgment

1.031mA

Positive

Normal

DC

PASS

1.024mA

Negative

Normal

DC

PASS

1.040mA

Positive

Power line

disconnected

DC

FAIL-U

1.019mA

Negative

Power line

disconnected

DC

PASS

1.013mA

Positive

Earth

disconnected

DC

FAIL-L

1.027mA

Negative

Earth

disconnected

DC

PASS

Save Data Command

MEMory:NUMBer

Query the data where the file no. is recorded and saved.

Query Syntax

MEMory:NUMBer?

Query Return

Returns the number of data files, ranging from 1~100.