REMOTE CONTROL

129

Manual Measurement Commands

CONFigure:CONDition

Sets or queries the EUT status when in manual measurement mode.
Note: The CONFigure:CONDition command or query can only be used in
manual measurement mode.

Syntax

CONFigure:CONDition
{NORMal|EARTh|POWersource|NAPPly|RAPPly|LLINe
|NLINe}

Query Syntax

CONFigure:CONDition?

Query Return

Returns the EUT status when in manual measurement
mode.
NORMal|EARTh|POWersource|NAPPly|RAPPly|LLINe|
NLINe

NORMal

Under normal conditions.

EARTh

Disconnected earth line.

POWersource

Disconnected live line.

NAPPly

Normal connection, positive phase,
110% voltage application.

RAPPly

Negative phase, 110% voltage
application.

LLINe

Application of voltage from the live
line. Normal live line connection.

NLINe

Application of voltage from the
neutral line. Normal neutral line
connection.

Example

CONFigure:CONDition NORMal
Set the leakage current test to normal conditions.

Query Example

CONFigure:CONDition ?
Return: NORMAL
The leakage current test is set to normal conditions.