Automatic measurement commands – GW Instek GLC-9000 User Manual User Manual

Page 134

background image

GLC-9000 User Manual

134

Example

CONFigure:POLarity NORMal
Sets the power supply to positive polarity.

Query Example

CONFigure:POLarity?
Return: NORMal
The polarity of the power supply is currently set to
positive.

Automatic Measurement Commands

CONFigure:AMITem

Configure or query all auto measurement settings. The settings must be
compatible with the measuring network, class and leakage mode, refer to
the list from the page 130. Any bits that are set to 1 indicate that the
corresponding mode/function is set.
Note: Automatic measurements cannot be made in manual
measurement mode.

Syntax

CONFigure: AMITem

Query Syntax

CONFigure:AMITem?

Query Return

Returns a 9-bit integer (1~512).

256

128

64

32

16

8

4

2

1

Bit8

bit7

bit6

bit5

bit4

bit3

bit2

bit1

bit0

NLINe

LLINe

Neg

phase

Pos

phase

RAPPly NAPPly EARTh

POWer-

source

NORMal

NORMal

Under normal conditions.

POWersource

Disconnected live line.

EARTh

Disconnected earth line.

NAPPly

Normal connection, positive phase,
110% voltage application.

RAPPly

Negative phase, 110% voltage
application.

LLINe

Application of voltage from the live
line. Normal live line connection.