GLC-9000 User Manual

16

Single Fault conditions include the following:
1. Earth disconnected (excluding earth leakage

current tests).

2. Disconnected neutral power line.
3. External equipment failure. (Patient leakage

current II, Patient Leakage Current III).

Polarity of the power supply can also be switched
to measure the leakage current under test. Thus
the polarity of the power supply should also be
taken into account.

Measurement
Methods

The diagram below shows how the probes, MD’s
and power supply are connected for different
leakage current tests.

Measuring Devices

Background

Leakage current (touch current) tests that measure
electrical equipment require a circuit network that can
simulate the impedance of a human body. The
impedance of a human body varies with the contact
points, area and the path of conduction. Thus the
circuit network used to simulate the impedance of a
human body varies with the type of test performed. As
such, the safety standards used to measure leakage
current also varies greatly. The circuit networks used
are known as measuring devices, or MD for short. MD