Leakage Current
Measurement
Modes

Earth leakage current
Enclosure to earth leakage current
Enclosure to enclosure leakage current
Enclosure and line leakage current
Patient Auxiliary current
Patient Leakage current I
Patient Leakage current II
Patient Leakage current III

Leakage Current
Type

DC, AC, AC+DC, ACpeak

Maximum
allowable
measurement
current

25mA (rms), 75mA (AC peak)

Leakage Current
Range

25mA (Max 25.00mA, Resolution 0.01mA)
5mA (Max 5.000mA, Resolution:0.001mA)
500uA (Max 500.0uA, Resolution:0.1uA)
50uA (Max 50.00uA, Resolution:0.01uA)

Range Switch

AUTO, HOLD

110% Voltage
Application

T3 output, internal 12k resistance protection