GLC-9000 User Manual

126

OFF: RC filter is OFF, pure 1k resistance is enabled.

Example

CONFigure:FILTer OFF
Turns the RC filter OFF for Measuring Network F.

Query Example

CONFigure:FILTer ?
Return: OFF
Returns the RC filter status. The RC filter is turned off.

Measurement Items

CONFigure:COMParator

Sets the current measuring mode’s upper and lower limit.

Syntax

CONFigure:COMParator {[[|,]]}

Query Syntax

CONFigure:COMParator?

Query Return

Returns the current measuring mode upper and lower
limit. The first parameter is the upper limit, the
second is the lower limit.

: Range: +4.000E-6 ~ +20.00E-03 (in Amps)

Example

CONFigure:COMParator +4.000E-03,+100.0E-06
Set the upper limit to 4mA and the lower limit to
100uA.

Query Example

CONFigure:COMParator?
Return: +4.000e-03,+1.000E-04
Returns an upper limit of 4mA and a lower limit of
100uA.

CONFigure:CURRent

Sets or queries the leakage current type.

Syntax

CONFigure:CURRent {ACDC|AC|DC|ACPeak}

Query Syntax

CONFigure:CURRent?