Patient leakage current i – GW Instek GLC-9000 User Manual User Manual

Page 53

background image

OPERATION

53

3. Press the reset button to

stop measurement.



GLC-9000

Warning

Turn the circuit breaker off before removing the EUT.

Ensure the power consumption of the EUT doesn’t
exceed the rated power limits.

Patient Leakage Current I

Network

Medical network, (MD-F) for the applied part.
Applicable for internal power supply and Class I
and Class II types.

Connection

As illustrated, ensure that the power source and
test leads are properly connected.

Internal Power
Supply

1. Connect the test leads to the T1 and T2

terminals.

2. Position the T1 test lead to an ungrounded

section of the enclosure.

3. Position the T2 test lead to the applied part of

the EUT.

Internal power supply