GLC-9000 User Manual

148

Auto Measurement test items,
range 1~512:

256

128

64

32

16

8

4

2

1

Bit8

bit7

bit6

bit5

bit4

bit3

bit2

bit1

bit0

NLINe

LLINe

Neg

phase

Pos

phase

RAPPly NAPPly EARTh

POWerso

urce

NORMal

Auto measurement measuring time

in seconds {NR1}

Auto measurement wait time in

seconds {NR1}

Note:

Measurement conditions and Power supply polarity
() are returned when a manual
measurement is set.
~ are the return values
when an automatic measurement is set.
0 is returned when there is no data.

Query Example

SYSTem:FILE? 6
Return:
Panel06,NO-6,CLA2,B,F,1,0,1,2,+2.500E-02,+0.400E-
02,0,0,123,1,1

The 6th file has the following configuration:

File name

Panel06

File No.

NO-6

Equipment class

CLASS-II

Network (medical)

B

Network,(circuit)

F network

Measurement mode

Enclosure - earth
leakage

Measurement method

automatic

Leakage current type

DC

Leakage current range

10.00mA