Remote control, Remote control settings, Emote control – GW Instek GLC-9000 User Manual User Manual
Page 111: Parity: none, Data bits: 8, Stop bits: 1, Address: 1~30
REMOTE CONTROL
111
R
EMOTE CONTROL
Remote Control Settings
Background
Remote control can be used with the USB, RS232 or
GPIB interfaces.
Interface
2
3
4
5
1
6
2
3
4
5
1
6
2
3
4
5
1
6
The rear panel RS-
232, USB(Device),
GPIB
interface
COM Port
Settings
Configure the PC comport setting as shown below:
Baud rate: 4800/9600/115200. See page 99 for
information on how to set the baud rate on the
GLC-9000.
Parity: None
Data bits: 8
Stop bits: 1
Data overflow control: None
Note
The USB settings are configured the same as the
RS232 settings.
GPIB Settings
Set the GPIB parameters as shown below:
Address: 1~30
Command Terminator: LF or CR+LF.
See page 99 for details on how to configure the GPIB
address on the GLC-9000.