REMOTE CONTROL

127

Query Return

Returns the leakage current type as a string:
ACDC|AC|DC|ACPeak

ACDC: AC+DC
AC: AC

DC: DC
ACPeak : AC PEAK

Example

CONFigure:CURRent DC
Set the leakage current type to DC.

Query Example

CONFigure:CURRent?
Return: DC
DC is the leakage current type.

CONFigure:RANGe

Sets or queries the current range.
Note: Leakage current ranges can only be selected when in manual
mode.

Syntax

CONFigure:RANGe{AUTO|HOLD1|HOLD2|HOLD3|
HOLD4}

Query Syntax

CONFigure:RANGe?

Query Return

Return value: Returns the current range as a string:
AUTO|HOLD1|HOLD2|HOLD3|HOLD4

When AC, DC, AC+DC leakage current is selected
(target):

AUTO

Automatic current range

HOLD1

25.00mA range

HOLD2

5.000mA range

HOLD3

500.0uA range

HOLD4

50.00uA range

When ACpeak leakage current is selected:

AUTO

Automatic current range

HOLD1

75.0mA file

HOLD2

10.00mA range

HOLD3

1.000mA range

HOLD4

500.0uA range