Output of time synchronization logs (fl0, sntp), Syntax, Example – Yokogawa Standalone MW100 User Manual
Page 72: Time synchronization client error messages
2-22
IM MW100-17E
Output of Time Synchronization Logs (FL0, SNTP)
• Output using the FL command.
• SNTP client logs are output. Up to 50 past logs are retained. If that number is
exceeded, old logs are overwritten with new ones.
Syntax
EACRLF
yy/mo/dd_hh:mm:ss_kkkkkkkkkk_mmm...mCRLF
.........................................
ENCRLF
yy
Year (00 to 99)
mo
Month (01 to 12)
dd
Day (01 to 31)
hh
Hour (00 to 23)
mm
Minute (00 to 59)
ss
Seconds (00 to 59)
kkk...k
Description
error : Error
report : Report of the execution results (no errors)
mmm...m
Error message (only upon error occurrence) or ADJUST (if no error)
_
Blank
Example
EA
99/05/11 12:20:30 error EBROKEN
99/05/11 12:30:00 report
EN
Time Synchronization Client Error Messages
Error Code
Description
Corrective Action
EDORMANT
Internal processing error
Servicing required.
ENOTACTIVE
Internal processing error
Servicing required.
EACTIVE
Internal processing error
Servicing required.
ELINK
Ethernet not powered
Check cable connections and power of
HUB
EHOSTNAME
SNTP server name not correct
Check whether destination server name
is set correctly.
ETCPIP
Internal processing error
Servicing required.
EINVAL
Internal processing error
Servicing required.
ESEND
Queries the time on the server
Check the IP address of the MW100.
failed
ETIMEOUT
Failed to receive response from
Check whether destination SNTP
server
server is running.
EBROKEN
Received time information
Check whether destination SNTP
corrupted
server is functioning correctly
2.2 ASCII Output