beautypg.com

Output of time synchronization logs (fl0, sntp), Syntax, Example – Yokogawa Standalone MW100 User Manual

Page 72: Time synchronization client error messages

background image

2-22

IM MW100-17E

Output of Time Synchronization Logs (FL0, SNTP)

• Output using the FL command.
• SNTP client logs are output. Up to 50 past logs are retained. If that number is

exceeded, old logs are overwritten with new ones.

Syntax

EACRLF
yy/mo/dd_hh:mm:ss_kkkkkkkkkk_mmm...mCRLF
.........................................
ENCRLF

yy

Year (00 to 99)

mo

Month (01 to 12)

dd

Day (01 to 31)

hh

Hour (00 to 23)

mm

Minute (00 to 59)

ss

Seconds (00 to 59)

kkk...k

Description

error : Error

report : Report of the execution results (no errors)

mmm...m

Error message (only upon error occurrence) or ADJUST (if no error)

_

Blank

Example

EA
99/05/11 12:20:30 error EBROKEN
99/05/11 12:30:00 report
EN

Time Synchronization Client Error Messages

Error Code

Description

Corrective Action

EDORMANT

Internal processing error

Servicing required.

ENOTACTIVE

Internal processing error

Servicing required.

EACTIVE

Internal processing error

Servicing required.

ELINK

Ethernet not powered

Check cable connections and power of

HUB

EHOSTNAME

SNTP server name not correct

Check whether destination server name

is set correctly.

ETCPIP

Internal processing error

Servicing required.

EINVAL

Internal processing error

Servicing required.

ESEND

Queries the time on the server

Check the IP address of the MW100.

failed

ETIMEOUT

Failed to receive response from

Check whether destination SNTP

server

server is running.

EBROKEN

Received time information

Check whether destination SNTP

corrupted

server is functioning correctly

2.2 ASCII Output