App index, Sf filter settings, Xb burnout settings – Yokogawa Standalone MW100 User Manual
Page 24: Xj reference junction compensation settings, Pf pulse input module chattering filter settings
1-17
IM MW100-17E
Commands
1
2
3
App
App
Index
When Setting Scaling
Setting
SRp1,p2,p3,p4,p5,p6,p7,p8,p9,p10
p1 Channel range (001 to 060)
p2 Input type
VOLT
DC voltage
TC
Thermocouple
RTD
Resistance temperature
detector
DI
Contact input
OHM
Resistance input
STR
Strain input
PULSE
Pulse
p3 Measuring range
p4 Lower limit of span
p5 Upper limit of span
p6 MATH type (SCALE)
p7 Scaling lower limit (-30000 to 30000)
p8 Scaling upper limit (-30000 to 30000)
p9 Decimal place (0 to 4)
p10 Unit (max 6 characters)
Query
SR[p1]?
Example
Convert the DC voltage measured on channel 2 to
DC current. Set the measuring range to 6V (high
resolution), lower limit of span to 1 V, upper limit of span
to 5 V, lower limit of scaling to 0.00A, and upper limit of
scaling to 5.00A.
SR002,VOLT,6VH,10000,50000,SCALE,0,500,
2,A
Explanation • Set the p3, p4, and p5 setting parameters according
to the table in section 1.3.
• Specify measurement channels on p1.
• For the characters that can be used for p10, see
user specified characters in section 1.3.
SF
Filter Settings
Setting
SFp1,p2
p1 Channel range (001 to 060)
p2 Filter coefficient (0, 5, 10, 20, 25,
40, 50, 100)
Query
SF[p1]?
Example
Set the filter coefficient for channel 2 to 5.
SF002,5
Explanation Specify measurement channels on p1. However, do not
set this on the channels of the digital input module.
XB
Burnout Settings
Setting
XBp1,p2
p1 Channel range (001 to 060)
p2 Burnout processing (OFF, UP, DOWN)
Query
XB[p1]?
Example
Set UP (+OVER) when burnout occurs on channel 1.
XB001,UP
Explanation Specify measurement channels that can accept TC
input for p1.
XJ
Reference Junction
Compensation Settings
Setting
When p2=INT
XJp1,p2
When p2=EXT
XJp1,p2,p3
For an Internal Compensation Circuit
Setting
XJp1,p2
p1 Channel range (001 to 060)
p2 Reference junction compensation
selection (INT)
Query
XJ[p1]?
Example
Set RJC on channel 1 to internal
compensation circuit.
XJ001,INT
Explanation Specify measurement channels that can accept TC
input for p1
For External RJC
Setting
XJp1,p2,p3
p1 Channel range (001 to 060)
p2 Reference junction compensation
selection (EXT)
p3 External RJC value (-20000 to 20000)
Query
XJ[p1]?
Example
Set RJC on channel 2 to external RJC, and the
compensation value to 0 µV.
XJ002,EXT,0
Explanation • Units of p3 are µV.
• Specify measurement channels that can accept TC
input for p1.
PF
Pulse Input Module Chattering
Filter Settings
Setting
PFp1,p2
p1 Channel range (001 to 060)
p2 Chattering filtering (OFF/ON)
Query
PF[p1]?
Example
Turn on the chattering filter on channel 23.
PF23,ON
Explanation For p1, specify channels included on the pulse input
module.
1.4 Setting Commands (System Related)