Output of operation error logs (fl0, error), Output of recording logs (fl0, record) – Yokogawa Standalone MW100 User Manual
Page 60
2-10
IM MW100-17E
Output of Operation Error Logs (FL0, ERROR)
• Output using the FL command.
• The operation error log (records) is output. Up to 50 past operation error logs are
retained. If that number is exceeded, old logs are overwritten with new ones.
• For information on the meanings of error codes, see the MW100 Data Acquisition Unit
User’s Manual (IM MW100-01E).
Syntax
EACRLF
yy/mo/dd_hh:mm:ss_nnn_mmm...mCRLF
................................................
ENCRLF
yy
Year (00 to 99)
mo
Month (01 to 12)
dd
Day (01 to 31)
hh
Hour (00 to 23)
mm
Minute (00 to 59)
ss
Seconds (00 to 59)
nnn
Error code (001 to 999)
mmm...m
Error message (up to 80 characters)
_
Blank
Example
EA
05/04/01 12:20:00 123 Range setting error.
05/04/01 12:30:00 456 Media access error.
EN
Output of Recording Logs (FL0, RECORD)
• Output using the FL command.
• The data acquisition log is output. Up to 250
*
past data acquisition logs are retained. If
that number is exceeded, old logs are overwritten with new ones.
• 1 line, 40 characters (fixed length).
• For information on the meanings of error codes, see the MW100 Data Acquisition Unit
User’s Manual (IM MW100-01E).
* Or “1021” on models with firmware version R3.02 or earlier.
Syntax
EACRLF
yy/mo/dd_hh:mm:ss_fffffff_mmm...mCRLF
................................................
ENCRLF
yy
Year (00 to 99)
mo
Month (01 to 12)
dd
Day (01 to 31)
hh
Hour (00 to 23)
mm
Minute (00 to 59)
ss
Seconds (00 to 59)
fffffff
Status ((Power, Card, Format, Record, Mode, Create,Trigger, Delete,
Recover, Error, Code, Time, Save)
mmm...m
Message
_
Blank
2.2 ASCII Output