beautypg.com

Multiple negative response, Ascii output, Binary output – Yokogawa Standalone MW100 User Manual

Page 52

background image

2-2

IM MW100-17E

2.1 Response Format

Multiple Negative Response

If one of several commands delimited with a sub-delimiter results in an error, a multiple
negative response is returned. The response consists of E2 followed by the error
position, error number, and CRLF, on one line.

• It is only output for the portion of the command that was erroneous. If multiple

commands resulted in errors, the responses are delimited with commas.

• The position of the error is indicated by a 1 for the first command, 2 for the second

command, and so on.

Syntax

E2_ee:nnnCRLF

(if only one error occurred)

E2_ee:nnn,ee:nnn,...,ee:nnnCRLF

(if multiple errors occurred)

ee

Error position (01 to 99)

nnn

Error number (001 to 999)

_

Blank

Example

E2 03:123,07:456

ASCII Output

For the format of each, see section 2.2.

Syntax

EACRLF
.......................CRLF
:
.......................CRLF
.......................CRLF
ENCRLF

Binary Output

Overview Diagram

'E'

'B'

CR

LF

Data length

Flag

Identifier

Header sum

ASCII data
Binary data

Data sum

Binary data

1 byte

EBCRLF

Indicates that data is in binary.

Data Length

The number of bytes equalling flag + indentifier + header sum + binary data + data

sum.

Header Sum

Sum value of data length + flag + identifier.

Binary Data

For the output format of each data, see section 2.3.