2 ascii output, Responses to queries, Ascii output -5 responses to queries -5 – Yokogawa Standalone MW100 User Manual
Page 55
2-5
IM MW100-17E
Response
1
2
3
App
App
Index
2.2 ASCII Output
ASCII data can be of the following types. The following explains the various formats.
• Response to query
• Output of the latest measured/computed data
• Output of the latest output data
• Output of setting data
• Output of the decimal place and units
• Output of logs and alarm summaries
• Output of status information
• Output of the relay status
• Output of the system recognition status
• Output of the media free space
• Output of file information
• Output of the strain input initial balance result
Note
In this section, CRLF means “carriage return/line feed.”
Responses to Queries
Following the format, the current command settings stored in ASCII are output.
Syntax
EACRLF
ttsss...sCRLF
....................
ENCRLF
Example 1: (SR?) parameter specified: No, relevant settings: Yes
EA
SR001,VOLT,2V
SR001,VOLT,2V
..................
SR060,VOLT,2V
EN
Example 2: (SR?) parameter specified: No, relevant settings: No
EA
EN
Example 3: (SR 001 ?) parameter specified: Yes
EA
SR001,VOLT,2V
EN