beautypg.com

Rtd (cont.), Resistance, Strain – Yokogawa Standalone MW100 User Manual

Page 15: Pulse, Analog output

background image

1-8

IM MW100-17E

• RTD (cont.)

Input Type

RTD

String

RTD

Measuring Range

Cu53 at 0°C

a=0.00426035

Cu100 at 0°C

a=0.00425

Pt25

Cu10 GE

(high resolution)

Cu10 L&N

(high resolution)

Cu10 WEED

(high resolution)

Cu10 BAILEY

(high resolution)

Pt100

[1]

(Anti noise) [2]

JPt100

[1]

(Anti noise) [2]

Pt100 GOST

Cu100 GOST

Cu50 GOST

Cu10 GOST

Pt500

Pt1000

String

Cu53

Cu100

Pt25

Cu10GEH

Cu10LNH

Cu10WEEDH

Cu10BAILEYH

Pt100-1R

Pt100-2R

JPt100-1R

JPt100-2R

Pt100G

Cu100G

Cu50G

Cu10G

Pt500

Pt1000

Setting Range for Lower/Upper Limit

of Span

–50.0 to 150.0 °C

–50.0 to 150.0 °C

–200.0 to 550.0 °C

–200.0 to 300.0 °C

–200.0 to 300.0 °C

–200.0 to 300.0 °C

–200.0 to 300.0 °C

–200.0 to 600.0 °C

–200.0 to 250.0 °C

–200.0 to 550.0 °C

–200.0 to 250.0 °C

–200.0 to 600.0 °C

–200.0 to 200.0 °C

–200.0 to 200.0 °C

–200.0 to 200.0 °C

–200.0 to 600.0 °C

–200.0 to 600.0 °C

Differential Computation

±200.0°C

±200.0°C

±750.0°C

±500.0°C

±500.0°C

±500.0°C

±500.0°C

±800.0°C

±450.0°C

±750.0°C

±450.0°C

±800.0°C

±400.0°C

±400.0°C

±400.0°C

±800.0°C

±800.0°C

The number in brackets for the setting range is the RTD measurement current (mA). If no number is

given, 1 mA, 2 mA, or both are supported.

• Resistance

Input Type

Resis-

tance

String

OHM

Measuring Range

20 Ω

200 Ω

2 kΩ

String

20ohm

200ohm

2000ohm

Setting Range for Lower/Upper Limit

of Span

0.000 to 20.000 Ω

0.00 to 200.00 Ω

0.0 to 2000.0 Ω

Differential Computation

±20.000 Ω

±200.00 Ω

±2000.0 Ω

• Strain

Input Type

Strain

String

STR

Measuring Range

2000 µSTR

20000 µSTR

200000 µSTR

String

2000uSTR

20000uSTR

200000uSTR

Setting Range for Lower/Upper

Limit of Span

–2000.0 to 2000.0 µSTR

–20000 to 20000 µSTR

–20000 to 20000 µSTR

Differential Computation

±2000.0 µSTR

±20000 µSTR

±200000 µSTR

The 200000 µSTR measuring range is 20000 (minimum resolution of 10 µSTR) on the

instrument’s setting/output data.

• Pulse

Input Type

Pulse

String

PULSE

Measuring Range

LEVEL

CONTACT

String

LEVEL

CONTACT

Setting Range for Lower/Upper Limit

of Span

0 to 30000

0 to 30000

Differential Computation

±30000

±30000

• Analog Output

Ouput Type

Volt

Current

String

AO

Output Range

Volt output (V)

Current output (mA)

String

10V

20mA

Setting Range for Lower/Upper Limit

of Span

–10.000 to 10.000 V

0.000 to 20.000 mA

Available output range

±11.000 V

0.000 to 22.000 mA

1.3 Parameters