beautypg.com

7 control commands, Control commands -38, Ds operation mode switching – Yokogawa Standalone MW100 User Manual

Page 45: Ex computation start/stop/reset/ clear, Ps recording start/stop, Mv setting file save/write, Mq measured, computed, and thinned file division, Ms message writing, Mp execution of manual sample, Ki main unit key operation

background image

1-38

IM MW100-17E

1.7 Control Commands

DS

Operation Mode Switching

Setting

DSp1
p1 Mode type
0

Measurement Mode

1

Setting Mode

Example

Sets the mode to Setting mode.
DS1

Explanation The operation mode cannot be changed during

recording or computation operations.

EX

Computation Start/Stop/Reset/

Clear

Setting

EXp1
p1 Operation
0

Computation start

1

Computation stop

2

Reset computation

3

Clear computation

Example

Starts computation.
EX0

Explanation • You can set this command on products with the

/M1 MATH function option, or when the pulse input
module is installed.

• Computation cannot be started or stopped during

recording.

• You cannot start computation if MATH is not turned

ON on at least one MATH channel using SO
command.

• Computation cannot be started if the MATH interval

is OFF, 10 ms, or 50 ms.

PS

Recording Start/Stop

Setting

PSp1
p1 Recording start/stop
0

Start

1

Stop

Example

Starts recording.
PS0

Explanation • When recording is started, measured data is saved

to the CF card.

• If the allowed size of a created measured value file,

computed value file, or thinned value file is exceeded
due to the combination of the recording channel,
recording interval, and recording data length,
recording will not be started.

• If the thinning recording interval is shorter than the

measurement or MATH interval, recording will not
start.

MV

Setting File Save/Write

Setting

MVp1,p2
p1 Operation
0

Save settings file

1

Load settings file

p2 File name (up to 8 characters)

Example

Loads settings file “SETFILE1.”
MV1,SETFILE1

Explanation • Specify the file name without the extension.

• For the characters that can be used for file names,

see “File Name and Foldr Name String” in section
1.3.

MQ

Measured, Computed, and

Thinned File Division

Setting

MQp1
p1 Operation
SAVE_M Execute division and saving

of measurement / computation
file

SAVE_T Execute division and saving

of thinned file

Example

Divides measurement files.
MQSAVE_M

Explanation After dividing a file, the next file division is enabled 10

minutes later.

MS

Message Writing

Setting

MSp1
p1 Message Number (0 to 5)

Example

Write message number 1.
MS1

MP

Execution of Manual Sample

Setting

MPp1
p1 Operation
SAMPLE Perform manual sample
DIVIDE Divide manual sample file

Example

Perform a manual sample.
MPSAMPLE

Explanation A new manual sample cannot be performed while a

manual sample file is being written (approximately 2
seconds).

KI

Main Unit Key Operation

Setting

KIp1
p1 Key type
START1 START key
START2 START key (equivalent to

holding down for 2 sec)

STOP1

STOP key

1.7 Control Commands