App index, Mh recording channel settings, Mw recording operation settings – Yokogawa Standalone MW100 User Manual
Page 34
1-27
IM MW100-17E
Commands
1
2
3
App
App
Index
Example
Create the folder automatically.
MDAUTO
Set the folder name to XXX100.
MDPARTIAL,XXX,100
Set the folder name to ZZZZZZ.
MDFREE,ZZZZZZ
Set the folder name to the date/time when the recording
is started.
MDDATE
Explanation • Folder type and format
Type
Naming
convention
Format
Handling of
duplicate folder
names
AUTO
DATAnnnn
DATA: Fixed
nnnn: Sequence number
0000 to 9999,
automatic update
Available
Sequence number
updating
PARTIAL xxxxnnnn
xxxx: Any 1 to 4
alphanumeric characters
nnnn: Sequence number
0000 to 9999,
automatic update
None
FREE
xxxxxxxx
xxxxxxxx: Any 1 to 8
alphanumeric characters
None
DATE
mddhhmmn
mddhhmm: Month, day,
hour, minute
m: 1 to 9, X (October),
Y (November), Z
(December)
dd: 01 to 31
hh: 00 to 23
mm: 00 to 59
n: Sequence number
(0 to 9, A to Z)
Available
Sequence number
updating
• The sequence number is managed by setting type.
You can set the starting sequence number for
the PARTIAL setting. You cannot set the starting
sequence number of AUTO and DATE settings.
• The folder name cannot be blank (only spaces).
Enter at least one character.
• If the number of characters of the folder name is 1
to 3 for the PARTIAL setting, the sequence number
is moved forward immediately after the folder
name. If the folder name is more than 4 characters,
characters after the 4th character are ignored. If the
folder name is set to DATA, consistency with the
auto setting is not maintained.
• In the FREE setting, the data is saved to the
same folder until the folder name is changed.
The recording log and alarm summary files are
overwritten when recorded to the same name. When
the recording stop action is set to Rotate, previous
old files are deleted.
• Commands can be sent or received with p2 and p3
attached regardless of the type (p1).
• For the character strings that can be used for the
folder name, see “File and Folder Name Strings” in
section 1.3.
MH
Recording Channel Settings
Setting
MHp1,p2,p3,p4
p1 Channel range (001 to 060 and A001 to
A300)
p2 Recording OFF/ON
p3 Thinning recording OFF/ON
p4 Manual sample recording OFF/ON
Query
MH[p1]?
Example
Set recording on channel 1 OFF, thinning recording ON,
and manual sample recording ON.
MH001,OFF,ON,ON
Explanation • Specify measurement channels on 001 to 060 of p1.
• If the /M1 MATH function option is not included,
MATH channels (A001 to A300) cannot be set
(including queries). However, if the pulse input
module is installed, MATH channels (A001 to A060)
can be set (including queries).
MW
Recording Operation Settings
Setting
When p2=OFF
MWp1,p2
When p2=ON and p3=DIRECT
MWp1,p2,p3,p4,p5
When p2=ON and p3=TRIGGER
MWp1,p2,p3,p4,p5,p6,p7
p1 Measurement group number (1 to 3)
p2 Recording OFF/ON
p3 Record start timing
DIRECT Record starts on record
start
TRIGGER Record starts on trigger
activation
p4 Recording stop action
SINGLE Recording stops after
creating one file
FULLSTOP Recording stops when file is
full
ROTATE Continue recording by
deleting the old data
p5 Recording interval [measurement/
times] (1, 2, 4, 5, 10)
p6 Recording data length
10M/20M/30M (units of minutes)
1H/2H/3H/4H/6H/8H/12H (units of
hours)
1D/2D/3D/5D/7D/10D (units of days)
p7 Pretrigger length[%] (0/10/20/30/40/5
0/60/70/80/90/100)
Query
MW[p1]?
Example
Turn recording of measurement group 1 ON, set the
record start timing to DIRECT, set the recording stop
action to SINGLE, and set the recording interval to 2.
MW1,ON,DIRECT,SINGLE,2
1.4 Setting Commands (System Related)