The main module has a two-digit 7-segment LED. The 7-segment LED displays the
system status.This section describes the displays on the 7-segment LED when errors
occur on the system and their corrective actions.
If servicing is necessary, or if the instrument is not operating correctly after performing
the corrective actions below, contact your nearest YOKOGAWA dealer.

Errors upon Startup

The left and right digits of the 7-segment LED display “b” and an error code, respectively.
The LED illuminates.

Display

Possible Problem

Corrective Action

b* (where * is The dipswitch settings are not correct.

Turn OFF the power, remove the CF card, turn ON all dip , switches

any character

and power up again. If the situation does not change servicing is

other than F).

required.

bF

The dipswitch settings are not correct.

Power up in setup reset mode.Turn OFF the power, turn ON all

dipswitches, and power up again. Since all settings such as the IP

address are initialized, reconfiguration is necessary.

System Errors

The left and right digits of the 7-segment LED display “F” and an error code, respectively.
The LED illuminates.

Display

Possible Problem

Corrective Action

F0

System ROM error.

Servicing required.

F1

SRAM error

Servicing required.

F2

EEPROM error

Servicing required.

F3

Error in the internal battery of the

Servicing required.

main module.

However, this error is also displayed immediately after the battery is

replaced. If this happens, power-cycle the MW100.

F4

Ethernet controller error

Servicing required.

F6

Web file load error

Servicing required.

FF

Error in writing unit information.

Servicing required.

Module Errors

The left and right digits of the 7-segment LED display are U and an error code,
respectively. The LED illuminates.
In the case of module errors, the error number and the corresponding module number
are displayed alternately as shown in the figure below.

Error number

Module number

Display

Possible Problem

Corrective Action

U0

Range information error.

Servicing required.

U1

Calibration value error.

Check the module’s installation status, then recalibrate the module.

If the error occurs even after recalibrating, servicing is required.

U2

Calibration reference voltage is not

Check whether the correct calibration reference voltage is or being

correct.

applied or whether the channel to which the voltage is applied is

(during calibration)

correct.

U3

Error in writing the calibration value.

Servicing required.

U4

The installed module cannot be used.

Replace the module with one that can be used.