Related instruments – Yokogawa RotaMASS 3-Series User Manual
Page 74

8. GENErAL SPECIFICATIONS
8-20
IM 01R04B05-00E-E 3rd edition July 30, 2010 -00
all Rights Reserved. Copyright © 2005, Rota Yokogawa
Advanced Concentration Measurement Options (others on request), recommended with option /K6:
Option
display
Components
Concentration
range
Temp. range
Source of concentration- / density table
/C01
°Brix
Sugar / Water
0 - 85 °Brix
0 - 80°C
PtB- Messages 100 5/90:
„the density of watery Saccarose solutions after the introduction of the international
temperature scale of 1990 (ItS1990)“ table 5
/C02
Wt%
naoH / Water
2 - 50 Wt%
0 - 100°C
D´ans-Lax, Handbook for chemists and physicists Vol.1, 3rd edition, 1967
/C03
Wt%
KoH / Water
0 - 60 Wt%
54 - 100°C
D´ans-Lax, Handbook for chemists and physicists Vol.1, 3rd edition, 1967
/C04
Wt%
nH
4
no
3
/ Water
1 - 50 Wt%
0 - 80°C
Source undefined
/C05
Wt%
nH
4
no
3
/ Water
20 - 70 Wt%
20 - 100°C
Source undefined
/C06
1
)
Wt%
HCl / Water
22 - 34 Wt%
20 - 40°C
D´ans-Lax, Handbook for chemists and physicists Vol.1, 3rd edition, 1967
/C07
Wt%
Hno
3
/ Water
50 - 67 Wt%
10 - 60°C
Source undefined
/C09
Wt%
H
2
o
2
/ Water
30 - 75 Wt%
4 - 44°C
Source undefined
/C10
Wt%
Ethylene Glycol / Water
10 - 50 Wt%
-20 - 40°C
Source undefined
/C11
Wt%
amylum = starch / Water
33 - 43 Wt%
35 - 45°C
Source undefined
/C12
Wt%
Methanol / Water
35 - 60 Wt%
0 - 40°C
Source undefined
/C20
VoL%
alcohol / Water
55 - 100 VoL%
10 - 40°C
Source undefined
/C21
°Brix
Sugar / Water
40 - 80 °Brix
75 - 100°C
Source undefined
1)
only with material HC
rELATEd INSTrUMENTS
the customer should prepare instrument maintenance tool, terminator, fieldbus power supply etc.
Safety barrier for version with intrinsically safe FF-output (option /KF4)
See web page www.yokogawa.com/fbs/Interoperability/fbs-accessories-en.htm