Yokogawa RotaMASS 3-Series User Manual
Page 43

5. CONFIGUrATION
5-19
IM 01R04B05-00E-E 3rd edition July 30 2010 -00
all Rights Reserved. Copyright © 2005, Rota Yokogawa
11) BI_dIrECTION
(relative Index 68)
Enables the RotaMaSS to measure the
flow in bi- or uni-direction mode. the sign
of the calculated flow rate values depends
on the flow direction.
Setting range: Bi-Direction; uni-Direction
Default: Bi-Direction
Parameters for local display
12) dISP_SELECT_1
(relative Index 50)
Selects the data to be displayed on the
first line of the LCD indicator, as follows
(Default : Massflow) :
1 = aI1 out = Massflow:
actual mass flow rate
2 = aI2 out = Volumeflow:
actual volume flow rate
3 = aI3 out = Density:
actual density value
4 = aI4 out = temperature:
actual temperature value
5 = aI5 out = Conc Meas:
actual calculated concentration value
6 = aI6 out = netflow:
actual net flow rate
7 = It1 out = It1: Integrator 1 totalized
value
8 = It2 out = It2: Integrator 2 totalized
value
9 = VELoCItY: actual medium´s velocity
6) VOLUME_FLOW_LOWCUT
(relative Index 32)
Sets the low cutoff volume flow rate level.
Setting range: Minimum flow rate 0 to 10%
of VoLuME_FLoW_VaLuE_RanGE.
Eu_100. the default value is 0 %.
unit: as selected in aI2.XD_SCaLE.
unItS_InDEX
Parameters for Tertiary Variable
7) dENSITy_VALUE_FTIME
(relative Index 35)
Defines the damping time constant for the
density to be input to the density calculation.
Setting range: 0.1 to 200 (seconds)
Default: 3 (seconds)
8) dENSITy_LOWCUT
(relative Index 36)
Sets the low cutoff density level.
Setting range: Minimum value 0 to 10%
of DEnSItY_VaLuE_RanGE.Eu_100.
the default value is 0 %.
unit: as selected in aI3.XD_SCaLE.
unItS_InDEX
Parameters for Quaternary Variable
9) TEMPErATUr_VALUE_FTIME
(relative Index 39)
Defines the damping time constant for the
temperature to be input to the temperature
calculation.
4) MASS_FLOW_LOWCUT
(relative Index 28)
Sets the low cutoff mass flow rate level.
Setting range: Minimum flow rate 0 to
10% of MaSS_FLoW_VaLuE_RanGE.
Eu_100. the default value is 0 %.
unit: as selected in aI1.XD_SCaLE.
unItS_InDEX
Parameters for Secondary Variable
5) VOLUME_FLOW_VALUE_FTIME
(relative Index 31)
Defines the damping time constant for the
volume flow rate to be input to the flow
rate calculation.
Setting range: 0.1 to 200 (seconds)
Default: 3 (seconds)
Setting range: 0.1 to 200 (seconds)
Default: 3 (seconds)
Parameters for flow direction
10) FLOW_dIrECTION
(relative Index 67)
Defines the direction of the flow and deter-
mines the sign of the calculated flow rate
values.
Setting range: Forward; Reverse
Default: Forward