Table A8.2 Transducer Status
Frequency Failure(E-01)
Error
Driving frequency out of range. (E-01). (f < Frequency low limit, or f > Frequency high limit)
[Remedy]: Check cable (remote) and detector.
Signal Failure(E-02)
Error
Phase difference out of range. (E-02).
(deltaPHI < Phase Difference low limit, or deltaPHI > Phase Difference high limit)
[Remedy]: Check cable (remote) and detector.
Sensor 1 Signal Failure(E-06)
Error
Sensor 1 signal too small. (E-06). (S1 < 7% of Drive amplitude for 3 min)
[Remedy]: Check cable (remote) and detector.
Sensor 2 Signal Failure(E-07)
Error
Sensor 2 signal too small. (E-07). (S2 < 7% of Drive amplitude for 3 min)
[Remedy]: Check cable (remote) and detector.
temp Sensor Failure(E-08)
Error
temperature sensor failed. (E-08). (t<-210degC, or t>450degC)
[Remedy]: Check cable (remote) and detector. Check temperature.
SoftDL Incomplete(E-09)
Error
Software download is incomplete. (E-09)
[Remedy]: Check the cables, power and RB Softdown Error(RB.SoFtDWn_ERRoR).
SoftDL Failure(E-10)
Error
Software download failed. (E-10)
[Remedy]: Check the download file and RB Softdown Error(RB.SoFtDWn_ERRoR).
abnormal Boot Process
Error
abnormal boot processing was detected at the time of starting.
[Remedy]: Check the cables, power and RB Softdown Error(RB.SoFtDWn_ERRoR).
Slug Detection(a-14)
alarm
Slug criterion is reached. (a-14)
[Remedy]: Reduce gas bubbles in medium.
Empty Pipe Detection(a-15)
alarm
Empty Pipe criterion is reached. (a-15)
[Remedy]: Fill pipe.
Corrosion Detection(a-16)
alarm
Corrosion criterion is reached. (a-16)
[Remedy]: Change detector.
Density Lower 0.3kg/l(W-01)
Warning
Density lower than 0.3kg/l. (W-01)
[Remedy]: Fill detector with fluid.
Fixed Dens Selected(W-02)
Warning
Reference density is enabled. (W-02)
[Remedy]: Set tB Density Fixed Value Selection (tB.DEnSItY_FIX_VaL_SEL) to inhibit.
Fixed temp Selected(W-03)
Warning
temperature fixed value is enabled. (W-03)
[Remedy]: Set tB temperature Fixed Value Selection (tB.tEMP_FIX_VaL_SELECt) to inhibit.
Fixed Mass Flow Selected(W-04)
Warning
Mass flow fixed value is enabled. (W-04)
[Remedy]: Set tB Mass Flow Fixed Value Selection (tB.MaSS_FLoW_FIX_VaL_SEL) to inhibit.
autozero Value out of Range(W-6)
Warning
autozero value is out of the internal detector depending range. (W-06)
[Remedy]: Stop flow during autozero. Check detector installation.
autozero Fluct out of Range(W-7)
Warning
autozero fluctuation is out of the internal detector depending range. (W-07)
[Remedy]: Stop flow during autozero. Check electrical installation, vibration and density.
PD/Freq Simulation active(W-08)
Warning
Fixed frequency or fixed phase difference is enabled. (W-08)
[Remedy]: Set tB Sensor Simulation (tB.SEnSoR_SIMuLatIon) to inhibit.
Figure A8.2 Transducer Status