beautypg.com

Electrical specifications, Remote cable rccy03 specification – Yokogawa RotaMASS 3-Series User Manual

Page 58

background image

8. GENErAL SPECIFICATIONS

8-4

IM 01R04B05-00E-E 3rd edition July 30, 2010 -00

all Rights Reserved. Copyright © 2005, Rota Yokogawa

ELECTrICAL SPECIFICATIONS

Power supply

- aC- type

: 90 V to 264 V

90 V to 250 V for Ex-type

- DC- type

: 20.5 V to 28.8 V

External circuit breaker rating : 5 a, 250 V (In the converter

no power switch is installed).

Fuse on Base Board :

- aC- type

: 2 a, t, breaking capacity 1500a

- DC- type

: 2 a, t, breaking capacity 1500a

display function

- up to 4 lines.

- 3 languages selectable (English, German, French)

- Instantaneous flow rate, density, temperature or totalized

flow can be displayed.

damping functions

adjustable from 0.1 seconds (63% response time) to 200

seconds, controls display and outputs.

Isolation resistance of converter

When surge arrestors are removed

- between power and ground terminal: 100 M

/ 500 V DC

- between power and I/o terminals : 20 M

/ 100 V DC

- between I/o terminals and ground : 20 M

/ 100 V DC

dielectric strength

When surge arrestors are removed

- between power and ground terminal : 1,500 V aC for 1 minute

Lightning Protection
arresters (2000 a) are inside converter for power supply
lines.

EMC
acc.

En 61326-1: 2006

En 61326-2-3: 2006

En 61000-3-2: 2006

En 61000-3-3: 1995+a1+a2

Supply Voltage of Communication Line :
- 9 V to 32 V DC for general purpose and flame-proof (/KF3) type
- 9 V to 24 V DC for intrinsic safe FF-output type (Entity model)
- 9 V to 17.5 V DC for intrinsic safe FF-output type (FISCo model)
no performance effect of power supply.

Output and Input Signal :
Digital communication signal based on FounDatIon
Fieldbus protocol.

rEMOTE CABLE rCCy03 SPECIFICATION

Li2Y(St)/CY 3x2 aWG24 + 1x3 aWG20
pair/triple shielded; pair/triple twisted; overall shielding
RCCY033/034 and RCCY031/032/KS1: flame propagation acc. IEC 60332-1.
table 6 : Cable specifications

Model code

Temperature

range

Wire gauge

resistance of

loop

Capacitance

wire/wire

Capacitance

wire/shield

Inductance

wire/wire

rCCy031/032

-50 to +70°C

aWG 24
aWG 20

190 Ω/km
70 Ω/km

157 nF/km
193 nF/km

249 nF/km
290 nF/km

0.60 mH/km
0.65 mH/km

rCCy031/032 /KS1

-50 to +70°C

aWG 24
aWG 20

190 Ω/km
70 Ω/km

157 nF/km
193 nF/km

249 nF/km
290 nF/km

0.60 mH/km
0.65 mH/km

rCCy033/034

-30 to +105°C

aWG 24
aWG 20

177 Ω/km
70 Ω/km

175 nF/km
145 nF/km

350 nF/km
290 nF/km

0.80 mH/km
0.70 mH/km

rCCy033/034 /KS1

-30 to +105°C

aWG 24
aWG 20

177 Ω/km
70 Ω/km

175 nF/km
145 nF/km

350 nF/km
290 nF/km

0.80 mH/km
0.70 mH/km

Condition of Communication Line :

- Supply voltage: 9 to 32 V DC

- Current draw: 15.0 ma (max)

Functional Specifications :

- Functional specifications for Fieldbus communication

conform to the standard specification (H1) of

FounDatIon Fieldbus.

Function Block :

- Four aI function blocks:

aI 1 monitors the mass flow rate

aI 2 monitors the volume flow rate

aI 3 monitors the density

aI 4 monitors the temperature

- two additional aI function blocks (with option /Cxx):

aI 5 monitors the measured concentration

aI 6 monitors the net flow rate

- one PID block (for a model with /LC1 option)

- two It function blocks (Integrator):

It 1 totalized mass-, volume- or net flow rate

It 2 totalized mass-, volume- or net flow rate

Update Period :

- Mass flow value: 100 ms

- Density, temperature: 100 ms

Function Block Execution Time

Block

name

Number Execution

time

Note

aI

6

< 30 ms

For mass flow, Volume flow,
Density, temperature,
Concentration measurement,
net flow

PID

1

< 50 ms

applicable when option /LC1
is selected

It

2

< 30 ms

For mass total, Volume total,
net total

Link Master function :

- Link Master (LM) function is supported.

See ‘Settings when shipped’ and ‘ordering information’.