8. GENErAL SPECIFICATIONS
8-1
IM 01R04B05-00E-E 3rd edition July 30, 2010 -00
all Rights Reserved. Copyright © 2005, Rota Yokogawa
8. GENErAL SPECIFICATIONS
Qnom is the water flow rate at about 1 bar pressure drop.
the flowmeter has an automatically low cut at 0.05% of
Qnom.
accuracy mass flow
:
Liquid :
± 0.1% of flow rate
± zero stability / flow rate *100%
(refer to table 2)
Gas (option /Ga) : ± 0.5% of flow rate
± zero stability / flow rate *100%
(refer to table 2)
accuracy volume flow
:
SQRt ( (mass flow error in %)² + (density error in %)²)
Please refer to sizing.
accuracy based on the frequency output includes the com-
bined effects of repeatability, linearity and hysteresis.
Repeatability for liquids: ± 0.05%
± (zero stability/2) / flow rate
*100%
Batch process :
above specified accuracy if the
batch process is >1 minute. For
shorter batch time (dt in s) the
accuracy decreases with the
square root of 60/dt
PErFOrMANCE SPECIFICATIONS
Model
- Remote detector RCCS30 to 33: 2 tubes, low flow design
- Remote detector RCCS34 to 39/XR : 2 tube design
- Remote field-mount converter RCCF31
- Integral type RCCt34 to 39/XR: 2 tube integral design
Fluid to be measured : Liquid, gas or slurries
Measurement items : Mass flow, density, temperature
and derived from these values:
concentration, volume flow and
net flow
Mass flow measurement
table 1: measuring range
Type
rCCS30 rCCS31 rCCS32 rCCS33
Qmax
t/h
0.1
0.3
0.6
1.5
Qnom
t/h
0.045
0.17
0.37
0.9
Type
rCCx34 rCCx36 rCCx38 rCCx39
rCCx39
/Ir
rCCx39
/Xr
Qmax
t/h
5
17
50
170
300
600
Qnom
t/h
2.7
10
32
100
250
500
table 2 : Zero Stability
Type rCCS30 rCCS31 rCCS32 rCCS33
kg/h
0.005
0.0085
0.019
0.045
Type rCCx34 rCCx36 rCCx38 rCCx39
rCCx39
/Ir
rCCx39
/Xr
kg/h
0.135
0.5
1.6
5
13
25
Pressure dependency
the stiffness of the RotaMaSS tubes is slightly line pressure
dependent. the static pressure effect of mass flow and den-
sity can be corrected by setting the static pressure manually via
menu.
table 3 : Static pressure effect on mass flow (not corrected)
Type
rCCS30 rCCS31 rCCS32
rCCS33
rCCx34
% of
rate
per bar
SS
----
----
----
----
0.00081
HC 0.00000
0.00012
0.00246
0.02094
0.00084
Type
rCCx36
rCCx38
rCCx39
rCCx39
/Ir
rCCx39
/Xr
% of
rate
per bar
SS
0.00346
0.00950
0.01058
0.02920
0.00740
HC
0.00336
0.00896
0.00808
0.01780
----
density measurement
adjustment with water and air at calibration temperature.
With option /K4 thermal stabilized.
For option /K6 see also “Special calibrations” on page 3.
Measuring range
: 0.3 kg/l to 5 kg/l (RCCx39,
RCCx39/IR and RCCx39/XR to 2 kg/l)
no density measurement for gas application
table 4: accuracy (at calibration conditions):
Type
Standard
Option /K4
Option /K6
RCCS30
0.008 g/cm³
------
------
RCCS31
0.004 g/cm³
0.001 g/cm³
------
RCCS32
0.004 g/cm³
0.001 g/cm³
0.0005 g/cm³
RCCS33
0.004 g/cm³
0.001 g/cm³
0.0005 g/cm³
RCCx34
0.003 g/cm³
0.001 g/cm³
0.0005 g/cm³
RCCx36
0.0022 g/cm³
0.001 g/cm³
0.0005 g/cm³
RCCx38
0.0015 g/cm³
0.001 g/cm³
0.0005 g/cm³
RCCx39
0.0015 g/cm³
0.001 g/cm³
0.0005 g/cm³
RCCx39/IR
0.0015 g/cm³
------
------
RCCx39/XR
0.0015 g/cm³
------
------
Repeatability:
- RCCS32-33, RCCx34-39/XR : ± 0.0005 g/cm³ (Std, /K4)
Static pressure effect:
Compensated if static pressure is set in the menu
Installation:
vertical, else correction term must be set in the software
Specification of high performance density measurement (option /K6):
ambient temp. range : -10°C to 50°C
Fluid temp. range : -50°C to 150°C
Minimum flow rate for specified accuracy:
- RCCx36 to RCCx39 : 700 kg/h
- RCCx34 : 140 kg/h
- RCCS33 : 90 kg/h
- RCCS32 : 37 kg/h
Maximum flow rate : Qnom
Repeatability : ± 0.0002 g/cm³
temperature measurement: ±0.5°C ±0.2% of reading
Density accuracy : ±0.0005 g/cm³ (liquids not aerated /
no gas in the liquid)
Process temperature influence :
0.000015 g/cm³ * abs(tfluid-20°C)
-1.0
-0.9
-0.8
-0.7
-0.6
-0.5
-0.4
-0.3
-0.2
-0.1
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
0
50
100
150
Flow in % of Qnom
Er
ro
r i
n
%
F10.EPS