25) VELOCITy_UNITS_INdEX
(relative Index 49)
Defines the unit of the medium velocity in
the tube. Velocity is zero if gas measure-
ment is selected.
Setting range: m/s, ft/s
Default: m/s
Parameters for 5th Variable (Option /Cxx)
26) CONCENTr_MEAS_VALUE_FTIME
(relative Index 42)
Defines the damping time constant for the
concentration meas value to be input to
the concentration calculation.
Setting range: 0.1 to 200 (seconds)
Default: 10 (seconds)
27) CONCENTr_MEAS_LOWCUT
(relative Index 43)
Sets the low cutoff concentration level.
Setting range: Minimum flow rate 0 to 10%
of ConCEntR_MEaS_VaLuE_RanGE.
Eu_100. the default value is 0 %.
unit: as selected in aI5.XD_SCaLE.
unItS_InDEX
Parameters for 6th Variable (Option /Cxx)
28) NET_FLOW_VALUE_FTIME
(relative Index 46)
Defines the damping time constant for the
net flow rate to be input to the flow rate
calculation.
Setting range: 0.1 to 200 (seconds)
Default: 3 (seconds)
29) NET_FLOW_VALUE_LOWCUT
(relative Index 47)
Sets the low cutoff net flow rate level.
Setting range: Minimum flow rate 0 to 10%
of nEt_FLoW_VaLuE_RanGE. Eu_100.
the default value is 0 %.
unit: as selected in aI6.XD_SCaLE.
unItS_InDEX