38 CLS User’s Guide
Installation
16 CLS Scaling Values
For linear mVdc, Vdc, and mAdc ranges, use 0.1% tolerance resistors.
Higher tolerances may cause significant errors. Correct any errors due
to resistor tolerance with the CLS’ built-in linear scaling. You can also
install other components (like capacitors) for signal conditioning; please
consult Watlow Anafaze for more information.
The next table shows scaling resistor values.
The next table shows the location of RC and RD on the analog input
board. (The analog input board is the upper board of the two-board set.)
A wire trace on the printed circuit board jumpers the RC position. When
you place a resistor in the RC position, cut the wire trace that connects
the two resistor terminals.
Input Range
RC
RD
All T/C, 0-60 mV DC
Jumper
0-10 mA DC
0-20 mA DC
Jumper
Jumper
6.0 ohms
3.0 ohms
0-100 mV
0-500 mV
499 ohms
5.49 Kohms
750 ohms
750 ohms
0-1 VDC
0-5 VDC
0-10 VDC
0-12 VDC
6.91 Kohms
39.2 Kohms
49.9 Kohms
84.5 Kohms
422.0 ohms
475.0 ohms
301.0 ohms
422.0 ohms
Loop #
RC
RD
Loop #
RC
RD
1
R58
R42
9
R57
R41
2
R56
R40
10
R55
R39
3
R54
R38
11
R53
R37
4
R52
R36
12
R51
R35
5
R50
R34
13
R49
R33
6
R48
R32
14
R47
R31
7
R46
R30
15
R45
R29
8
R44
R28
16
R43
R27
Analog
IN +
RD
CLS Measurement
Circuitry
Analog
Input
Terminals
RC
Common
IN +