Entering System Configuration Mode
In order to make any changes to unit configuration, you must first power on the unit in System
Configuration mode.
Procedure
Follow these steps to power on the unit in System Configuration mode:
Step
Action
1.
Turn the selector switch to OFF, and wait 10
seconds.
Simultaneously press and hold the right-most and
left-most softkeys (below the monitor) while
turning the selector switch to MONITOR (ON for
AED units), and hold the softkeys for
approximately 4 seconds.
The unit powers on in System Configuration
mode.
2.
Enter the passcode using the appropriate
softkeys.
Note: When you enter System Configuration for
the first time, enter the default pass code
of 00000000.
To change a digit, use the Delete softkey to
highlight the entry and re-enter the digit.
When all 8 digits have been entered, the unit
automatically advances to the Select Language
screen.
3.
Use the Prev Value and Next Value softkeys to
toggle the highlighted area between the language
options.
4.
Press the Enter softkey to select the highlighted
language.
All subsequent configuration menus will be in the
selected language.
The unit automatically advances to the Main
Configuration menu.
0
3
Delete
2
1
System Configuration
Enter Configuration Passcode
or Power Off to Exi
t
? ? ? ? ? ? ? ?
0
3
Delete
2
1
System Configuration
Enter Configuration Passcode
or Power Off to Exi
t
0 0 0 ? ? ? ? ?
Enter
Exit
Config
Next
Value
Prev
Value
System Configuration
Select Language
ENGLISH
FRANCAIS
Config
Change
Config
Exit
Config
Read
From
Card
Restore
Defaults
System Configuration