terminates any automatic analysis/re-analysis sequencing. If the ANALYZE button is pressed,
the unit analyzes only once.
Values are as follows:
Stacked Shocks
The parameter specifies the number of stacked shocks that must be administered prior to the
activation of a CPR interval and works in conjunction with the Auto Analyze parameter.
Values are as follows:
Display DO CPR Message
This parameter allows you to specify whether the unit displays the configured CPR prompt
(described below) for the configured CPR time interval after each set of stacked shocks, even if
a perfusing rhythm is present. The unit behaves differently depending on this particular
parameter value.
•
When set to No (the Manual mode default), the unit behaves as follows:
—
At the end of any analysis with an outcome of no shock advised, the unit displays the NO
SHOCK ADVISED message. The unit then remains idle for the configured CPR interval
duration. After the CPR interval, if the ANALYZE button has not been pressed, the
PRESS ANALYZE message is displayed and remains until a new analysis is started or the
mode is changed.
—
At the end of a stacked shock sequence (assuming the final shock is delivered) and when
the optional 10 second CHECK PULSE message disappears, the CPR interval begins.
The unit remains idle for the duration of the CPR interval. After the CPR interval, if the
ANALYZE button has not been pressed, the PRESS ANALYZE prompt is displayed and
remains until a new analysis is started or the mode is changed.
—
If the defibrillator disarms itself after it was charged (because the Defib Ready time
elapsed before a shock was delivered), the unit displays the PRESS ANALYZE prompt
until the ANALYZE button is pressed.
•
When set to Yes, the unit behaves as follows:
—
At the end of any analysis with an outcome of no shock advised, the unit displays the NO
SHOCK ADVISED message for 10 seconds, after which the message is cleared and the
configured CPR prompt is displayed for the duration of the CPR interval. After the CPR
interval, the PRESS ANALYZE prompt is displayed until a new analysis is started or a
mode change occurs. You can start a new analysis during the CPR interval by pressing
the ANALYZE button.
Possible Values:
Yes, No, Semi-Auto Only
Default:
Manual: No, AED: Semi-Auto Only
Possible Values:
1, 3
Default:
1