Alarm parameters, Alarms active at power up, Alarms limits at power up – ZOLL E Series Monitor Defibrillator Rev M User Manual
Page 45

Alarm Parameters
9650-1201-01 Rev. M
39
Alarm Parameters
Alarm parameters are attributes associated with Alarm states and thresholds for the E Series
unit and include:
•
Alarms Active at Power Up
•
Alarms Limits at Power Up
Alarms Active at Power Up
This parameter allows you to specify whether alarms are automatically active at power on. This
setting applies to all alarms enabled by the Alarm Limits At Power Up parameter.
Values are as follows:
Alarms Limits at Power Up
This option allows you to enable or disable, and set high and low threshold values for each
alarm on the E Series unit. Possible values and defaults for each alarm are as follows:
Possible Values:
Yes, No
Default:
No
Alarm
Maximum Default/Values
Minimum Default/Values
Heart Rate
Default: 150 BPM
Possible Values: 60 - 280 beats per
minute (BPM) inclusive in 5 BPM
increments
Default: 30 BPM
Possible Values: 30 - 100 beats per
minute (BPM) inclusive in 5 BPM
increments
SpO
2
Default: 100%
Possible Values: 72 - 100%
Default: 85%
Possible Values: 70 - 98%
SpCO
Default: 10%
Possible Values: 2 - 100%
Default: 0%
Possible Values: 0 - 99%
SpMet
Default: 3%
Possible Values: 1 - 100% as
follows:
0.1% increments from 1 - 2%;
0.5% increments from 2 - 10%;
1% increments from 10 - 100%
Default: 0%
Possible Values: 0 - 99% as follows:
0.1% increments from 0 - 2%;
0.5% increments from 2 - 10%;
1% increments from 10 - 99%
Respiration Rate
Default: 120 breaths/min.
Possible Values: 5 - 150 breaths/
min. in increments of 5
Default: 5 breaths/min.
Possible Values: 0 - 100 breaths/
min. in increments of 5
EtCO
2
Default: 55 mmHg or 7.3 kPa
Possible Values: In mmHg: 55 - 100
mmHg in increments of 5
In kPa: 8.0, 8.6, 9.3, 10.0, 10.6, 11.3,
12.0, 12.6, 13.3
Default: 25 mmHg or 3.3 kPa
Possible Values: In mmHg: 0 - 50
mmHg in increments of 5
In kPa: 0.0, 0.6, 1.3, 2.0, 2.6, 3.3,
4.0, 4.6, 5.3, 6.0, 6.6, 7.3