beautypg.com

Etco, Parameters, Default etco – ZOLL E Series Monitor Defibrillator Rev M User Manual

Page 41: Units, Displayed zoom level, Comp setting

background image

EtCO2 Parameters

9650-1201-01 Rev. M

35

EtCO

2

Parameters

EtCO

2

configuration parameters are available for update only if the EtCO

2

monitoring option is

installed on the unit and include:

Default EtCO2 Units

Displayed Zoom Level

Default EtCO2 Comp Setting

Power Save Mode at Power Up

EtCO

2

Average On Power Up

Default EtCO

2

Units

This parameter allows you to set the EtCO

2

unit of measurement.

Values are as follows:

Displayed Zoom Level

This parameter allows you to set the default scale setting for the displayed capnographic
waveform. It does not affect the printed waveform or the waveform data stored on the data
card.

Values are as follows:

Default EtCO

2

Comp Setting

This parameter allows you to set the default compensation for CO

2

.

Values are as follows:

Possible Values:

mmHg (millimeters of mercury), kPa (kilopascals),% (percent)

Default:

mmHg

Possible Values:

mmHg: 0-12.5, 0-25, 0-50, 0-75, 0-100
kPa: 0-1.7, 0-3.3, 0-6.6, 0-10.0, 0-13.3

Default:

mmHg: 0-50
kPa: 0-6.6

Possible Values:

None, O2, N2O, O2 & N2O

Default:

None