Etco, Parameters, Default etco – ZOLL E Series Monitor Defibrillator Rev M User Manual
Page 41: Units, Displayed zoom level, Comp setting

EtCO2 Parameters
9650-1201-01 Rev. M
35
EtCO
2
Parameters
EtCO
2
configuration parameters are available for update only if the EtCO
2
monitoring option is
installed on the unit and include:
•
Default EtCO2 Units
•
Displayed Zoom Level
•
Default EtCO2 Comp Setting
•
Power Save Mode at Power Up
•
EtCO
2
Average On Power Up
Default EtCO
2
Units
This parameter allows you to set the EtCO
2
unit of measurement.
Values are as follows:
Displayed Zoom Level
This parameter allows you to set the default scale setting for the displayed capnographic
waveform. It does not affect the printed waveform or the waveform data stored on the data
card.
Values are as follows:
Default EtCO
2
Comp Setting
This parameter allows you to set the default compensation for CO
2
.
Values are as follows:
Possible Values:
mmHg (millimeters of mercury), kPa (kilopascals),% (percent)
Default:
mmHg
Possible Values:
mmHg: 0-12.5, 0-25, 0-50, 0-75, 0-100
kPa: 0-1.7, 0-3.3, 0-6.6, 0-10.0, 0-13.3
Default:
mmHg: 0-50
kPa: 0-6.6
Possible Values:
None, O2, N2O, O2 & N2O
Default:
None