Advisory Defibrillation Parameters

The Advisory Defibrillation parameters control functions related to performing defibrillation
with an E Series unit in semiautomatic and advisory mode and include:

Energy Level: Shock 1

Energy Level: Shock 2

Energy Level: Shock 3

Auto Charge in Advisory Mode

Auto Analyze

Stacked Shocks

Display DO CPR Message

CPR Message After Shock

CPR Message After No Shock Advised

Duration of CPR Interval

Check Pulse

Restart Analysis After CPR

Check Patient Prompt

Enable Voice in Advisory Mode

Continual Check Patient Audio Prompt

Enable SCE

Post SCE Result on Display

Energy Level: Shock 1

This parameter allows you to select the energy level used for the first shock delivered by the
defibrillator when the unit is:

in semi-automatic or advisory mode and using pads

in Manual mode, using external paddles or pads, and Basic Auto Energy Escalation is set to
Yes

Energy Level: Shock 1 is the energy level to which the unit defaults when power is turned on
with external paddles or pads connected.

Values are as follows:

Energy Level: Shock 2

This parameter allows you to select the energy level used for the second shock delivered by the
defibrillator when the unit is:

in semi-automatic or advisory mode and using pads

in Manual mode, using external paddles or pads, and Basic Auto Energy Escalation is set to
Yes

Possible Values:

1-10, 15, 20, 30, 50, 70, 85, 100, 120, 150, and 200 J

Default:

120 J