Chapter 2
Configuration Parameters
16
9650-1201-01 Rev. M
Advisory Defibrillation Parameters
The Advisory Defibrillation parameters control functions related to performing defibrillation
with an E Series unit in semiautomatic and advisory mode and include:
•
Energy Level: Shock 1
•
Energy Level: Shock 2
•
Energy Level: Shock 3
•
Auto Charge in Advisory Mode
•
Auto Analyze
•
Stacked Shocks
•
Display DO CPR Message
•
CPR Message After Shock
•
CPR Message After No Shock Advised
•
Duration of CPR Interval
•
Check Pulse
•
Restart Analysis After CPR
•
Check Patient Prompt
•
Enable Voice in Advisory Mode
•
Continual Check Patient Audio Prompt
•
Enable SCE
•
Post SCE Result on Display
Energy Level: Shock 1
This parameter allows you to select the energy level used for the first shock delivered by the
defibrillator when the unit is:
•
in semi-automatic or advisory mode and using pads
•
in Manual mode, using external paddles or pads, and Basic Auto Energy Escalation is set to
Yes
Energy Level: Shock 1 is the energy level to which the unit defaults when power is turned on
with external paddles or pads connected.
Values are as follows:
Energy Level: Shock 2
This parameter allows you to select the energy level used for the second shock delivered by the
defibrillator when the unit is:
•
in semi-automatic or advisory mode and using pads
•
in Manual mode, using external paddles or pads, and Basic Auto Energy Escalation is set to
Yes
Possible Values:
1-10, 15, 20, 30, 50, 70, 85, 100, 120, 150, and 200 J
Default:
120 J