Chapter 2
Configuration Parameters
20
9650-1201-01 Rev. M
Check Pulse
This parameter works in conjunction with the Auto Analyze parameter and determines whether
the unit displays the prompt CHECK PULSE for 10 seconds immediately after delivery of the
final shock in a stacked shock sequence, after each No Shock Advised analysis result, or both.
During this 10-second time interval, the CHECK PATIENT and PRESS ANALYZE prompts are
suppressed on AED units in semi-auto mode, although background analysis is active. After
CHECK PULSE clears, the CPR interval begins.
When set to No, the unit immediately begins the CPR interval following delivery of the final
shock.
Values are as follows:
Restart Analysis After CPR
When this parameter is set to Yes, the analysis starts automatically after the CPR interval
following each shock. Analyses do not automatically restart after a CPR interval following No
Shock Advised results. When this parameter is set to No, the PRESS ANALYZE message is
displayed, but the analysis does not restart.
Values are as follows:
Check Patient Prompt
This parameter allows you to change the display message and the voice prompt issued when a
shockable rhythm has been detected by background ECG analysis and Heart Rate Alarms are
enabled or the unit is operating in Semi-Automatic mode.
Values are as follows:
Possible Values:
Yes, No, After No Shock Only, After Shock Only
Default:
No
Possible Values:
Yes, No
Default:
Yes
Possible Values:
CHECK PATIENT, PRESS ANALYZE
Default:
CHECK PATIENT