Nibp parameters, Nibp units, Trigger nibp meas on bp alarm – ZOLL E Series Monitor Defibrillator Rev M User Manual
Page 43: Trigger nibp meas on hr alarm

NIBP Parameters
9650-1201-01 Rev. M
37
NIBP Parameters
NIBP configuration parameters are available for update only if the NIBP monitoring option is
installed on the unit, and include:
•
NIBP Units
•
Trigger NIBP Meas on BP Alarm
•
Trigger NIBP Meas on HR Alarm
•
NIBP Auto Interval Default
•
NIBP First Inflate Pressure Default
•
NIBP Button Press and Hold
NIBP Units
This parameter allows you to set the NIBP unit of measurement.
Values are as follows:
Trigger NIBP Meas on BP Alarm
This parameter allows you to specify whether or not the E Series unit initiates a single
additional blood pressure measurement when any of the NIBP alarms trigger.
Values are as follows:
Trigger NIBP Meas on HR Alarm
This parameter allows you to specify whether or not the E Series unit initiates a single blood
pressure measurement when the heart rate/pulse rate alarm triggers.
Values are as follows:
Possible Values:
mmHg (millimeters of mercury), kPa (kilopascals)
Default:
mmHg
Possible Values:
Yes, No
Default:
Yes
Possible Values:
Yes, No
Default:
No