Nibp parameters, Nibp units, Trigger nibp meas on bp alarm – ZOLL E Series Monitor Defibrillator Rev M User Manual

Page 43: Trigger nibp meas on hr alarm

background image

NIBP Parameters

9650-1201-01 Rev. M

37

NIBP Parameters

NIBP configuration parameters are available for update only if the NIBP monitoring option is
installed on the unit, and include:

NIBP Units

Trigger NIBP Meas on BP Alarm

Trigger NIBP Meas on HR Alarm

NIBP Auto Interval Default

NIBP First Inflate Pressure Default

NIBP Button Press and Hold

NIBP Units

This parameter allows you to set the NIBP unit of measurement.

Values are as follows:

Trigger NIBP Meas on BP Alarm

This parameter allows you to specify whether or not the E Series unit initiates a single
additional blood pressure measurement when any of the NIBP alarms trigger.

Values are as follows:

Trigger NIBP Meas on HR Alarm

This parameter allows you to specify whether or not the E Series unit initiates a single blood
pressure measurement when the heart rate/pulse rate alarm triggers.

Values are as follows:

Possible Values:

mmHg (millimeters of mercury), kPa (kilopascals)

Default:

mmHg

Possible Values:

Yes, No

Default:

Yes

Possible Values:

Yes, No

Default:

No